The BedJet VET includes the most robust set of redundant safety features of any forced air animal

warmer currently on the market to prevent unintentional overheating. These features include:

  • Easily set Auto‐shutoff times from 30 minutes to 4 hours
  • Dual temperature sensors with microprocessor monitoring for error check — sensors are averaged for temperature readings and must agree with each other or the system will log a fault and shut down
  • Dual microprocessors with error monitoring and watchdog functions
  • 3‐stage microprocessor monitored overheat protection that will shut down heat under the following out of specification over‐temperature situations:
    • Air exhaust measured at > 46°C for 120 seconds
    • Air exhaust measured at > 53°C for 12 seconds
    • Air exhaust measured at > 55°C for 1 second
  • Mechanical thermostatic overheat switch located in air exhaust that will shut off